RADIATION DOSE TO PATIENTS UNDERGOING CHEST RADIOGRAPH AND BRAIN CT SCAN IN THE RADIOLOGY DEPARTMENT OF FEDERAL MEDICAL CENTRE KATSINA
Abstract
This study presents a comparative analysis of radiation dose indices across two diagnostic
modalities: conventional projection radiography and computed tomography (CT), utilizing data
from two distinct patient cohorts (N=20 each). The primary objective was to evaluate the
relationship between technical exposure parameters—specifically tube voltage (kVp), tube
current-time product (mAs), and exposure time—and the resulting radiation dose markers,
including estimated dose for radiography and CTDIvol and DLP for CT. For the radiography
cohort, patient ages ranged from 29 to 55 years. Analysis revealed that dose levels were highly
sensitive to fluctuations in kVp (70–76 kVp) and mAs (9–12 mAs), with estimated doses remaining
within a low-level safety margin of 0.17 to 0.22 mGy. The findings suggest a high level of protocol
standardization in conventional imaging, where adjustments are minimal and primarily driven by
anatomical requirements while maintaining a narrow dose variance of only 0.05 mGy. In contrast,
the CT cohort (ages 29–54) demonstrated the implementation of a fixed-voltage protocol (120
kVp), with dose modulation achieved exclusively through mAs variations (225–265 mAs). This
resulted in CTDIvol values ranging from 53 to 61 mGy and DLP values between 910 and 1040
mGy·cm. The direct correlation between mAs and dose indices in CT highlights the clinical
significance of current modulation as a tool for dose optimization. While CT doses are naturally
orders of magnitude higher than those in projection radiography, both modalities exhibited a trend
toward precision and consistency. The results underscore the importance of technical parameter
selection in adhering to the ALARA (As Low As Reasonably Achievable) principle. The uniformity
in kVp for CT scans suggests that peak energy is often standardized for specific diagnostic tasks,
while mAs serves as the primary variable for adjusting to individual patient characteristics.
Conversely, radiography protocols involve more frequent adjustments to both voltage and current,
though the absolute dose variation remains small.